SOURCE SKIN DISTANCE DEPENDENCE OF RESPONSE OF IN VIVO EXIT DOSE DIODE DETECTORS USED IN

  • 1 “Eqrem Cabej” University, Gjirokaster, Albania
  • 2 Hygeia Hospital Tirana, Albania
  • 3 Polytechnic University, Tirana, Albania

Abstract

Linear Accelerators and Cobalt units are common source of radiation for external radiotherapy. In Vivo Dosimetry (IVD) is a real time quality control set of methods used in cancer treatment clinics to assure that dose is delivered as prescribed to tumor, while surrounding healthy tissue is spared. Silicon Diode detectors are among most popular detectors used for IVD. In certain clinical conditions, both entrance and exit dose measurements are needed, for quality insurance, to calculate the dose received by tissue or organ inside patient’s body. Response of silicon diode detectors depends on multiple factors therefore prior of use in IVD each detector needs to be calibrated and corresponding factors of correction need to be calculated. Correction factors are needed to calculate dose for different field sizes, angle between beam axis and couch axis, skin to source distance (SSD), temperature etc. In this work we present our calculation of exit dose calibration factors for six silicon diode detectors made by PTW Frieburg Germany. Diodes are calibrated for exit dose under either 6 MVor 18 MV beam of photons as designated by manufacturer. Further we study dependence of response of individual diodes on distance from source and calculate corresponding correction factors. It was found that exit dose SSD dependence is pronounced and correction factors are necessary for exit dose IVD with diode detector in clinics.